Revision A
Dash 3000/4000 Patient Monitor
6-9
2000966-339
Configuration:
Boot Code Selections
Boot Code Selections
Set Defib Sync Voltage and Pulse Width
The monitor controls the analog out signal used to trigger a defibrillator.
Refer to the defibrillator manufacturer’s manual for the required pulse
amplitude and duration.
Use the Boot Code Service Menu to configure or change the Marker Out
signal of the
DEFIB SYNC
connector.
1. Activate the Boot Code.
a. Hold down
NBP GO/STOP
and
ZERO ALL
.
b. Press and release the
Trim Knob
control.
c. Keep holding
NBP GO/STOP
and
ZERO ALL
until the Boot
Code information appears on the display.
2. Select
SERVICE MENU
.
3. Select
SET CONFIGURATION
menu option.
4. In the Configuration Menu, select:
1 Defib Sync Voltage:
and choose
5V
or
12V
amplitude.
2 Defib Sync Pulse Width:
and choose
10 ms
or
100 ms
for pulse
duration.
5. Select
Exit
and wait for the monitor to display the message,
“Writing
settings to EEPROM...”
6. Manually reboot the monitor to implement this change.
Set Line Frequency
Use the Boot Code Service Menu to configure or change the monitor line
frequency to 50 or 60 Hz. The default is 60 Hz.
1. Activate the Boot Code.
a. Hold down
NBP GO/STOP
and
ZERO ALL
.
b. Press and release the
Trim Knob
control.
c. Keep holding
NBP GO/STOP
and
ZERO ALL
until the Boot
Code information appears on the display.
2. Select
SERVICE MENU
.
3. Select
SET CONFIGURATION
menu option.
4. Select
Line Frequency
then choose 50 Hz or 60 Hz line frequency.
5. Select
Exit
and wait for the monitor to display the message,
“Writing
settings to EEPROM...”
6. Manually reboot the monitor to implement this change.
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......