5-6
CARESCAPE V100 Vital Signs Monitor
2048723-002C
Troubleshooting:
Alarm code interpretation
E61
Temp
Temp probe broken (Alaris only)
No
Replace temperature probe
E63
Temp
Temp probe disconnected (Alaris only)
Yes
Check for correct probe
E66
Temp
Temp probe too hot (Alaris only)
Yes
Replace probe
E--
Temp
Temp scanner error (Exergen only)
Yes
Check Exergen battery, and
check for valid temperature
value. Also check the
scanner’s display for
Exergen error information.
Refer to the
section
for more information.
E80
NIBP
Overpressure
Yes
Excess cuff pressure. Check
for pinched or occluded
internal tubing.
E82
NIBP
Excess air in cuff
Yes
Determination cannot be
made due to an excess
amount of air in the cuff.
E83
NIBP
NIBP pump timeout
Yes
Leak in cuff or o-ring in
hose
Internal leak in tubing or
pneumatic valve
Pump not turning on
Check or replace hose or
cuff.
E84
NIBP
NIBP total timeout
Yes
Length of time has
exceeded 2 minutes for an
adult/pediatric
determination or 85
seconds for a neonate
determination.
E85
NIBP
NIBP level timeout
Yes
Remained at one cuff
pressure level for more than
1 minute.
E89
NIBP
NIBP no determination
Yes
Cuff placement on patient
*
Acknowledging an alarm by pressing the
Silence
button, cancels and resets the alarm condition.
Error code or
problem
Source
Definition
Can be
acknowledged
(silenced)?*
Probable source
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......