2048723-002C
CARESCAPE V100 Vital Signs Monitor
4-5
Maintenance:
Cleaning
Cleaning and disinfecting blood pressure cuffs
General
The cuff must be thoroughly cleaned with the specified detergent before reuse.
The additional use of household bleach as described below provides at least
intermediate-level disinfection.
NOTE
Never autoclave or steam clean the monitor, cuffs, or accessories.
Apply cuff hose caps before cleaning. See
for the orderable part numbers of cuff hose caps.
While this procedure is adequate for cleaning/disinfection, it may not
remove all stains.
Do
not
immerse hoses.
Do
not
immerse cuffs without prior application of the cuff hose caps.
Materials
Enzymatic detergent such as ENZOL
®
enzymatic detergent (US) or
Cidezyme
®
enzymatic detergent (UK)
Distilled water
10% solution of household bleach (5.25% sodium hypochlorite) in distilled
water
Soft cloths and soft-bristled brushes
Spray bottles
Procedure
1. Prepare the enzymatic detergent according to the manufacturer’s
instructions and the 10% bleach solution, in separate spray bottles.
2. Spray the detergent liberally on the cuff. If the material is dried on, allow the
detergent to set for 1 minute. For soil on the soft part of the closure or the
cuff itself, wipe the material off with a soft cloth. For persistent
contamination on the soft part of the closure, use a soft-bristled brush to
loosen particles. Rinse with copious amounts of distilled water. Repeat until
no visible contamination remains. For soil on the hook part of the closure,
use a soft-bristled brush to remove the material, and rinse with copious
amounts of distilled water. Repeat until no visible contamination remains.
3. Spray the 10% bleach solution on the affected area until the area is
saturated. Allow the solution to set for 5 minutes.
4. Wipe away any excess solution and rinse the cuff again with distilled water.
Allow 2 hours for drying.
The user has the responsibility to validate any deviations from the
recommended method of cleaning and disinfection.
For additional information on infection control procedures, contact GE Technical
Support.
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......