![GE CareScape V100 Скачать руководство пользователя страница 77](http://html.mh-extra.com/html/ge/carescape-v100/carescape-v100_service-manual_128562077.webp)
2048723-002C
CARESCAPE V100 Vital Signs Monitor
4-19
Maintenance:
Calibration procedures and tests
Alaris temperature calibration verification
NOTES
There are no user-performed maintenance or calibration procedures for
Alaris thermometry.
To verify the accuracy of the Alaris temperature parameter with a specific
temperature probe, take measurements at varying temperatures in
continuous (monitor) mode using a water bath and a high-precision
thermometer as a reference instrument.
1. Turn the monitor off. Make sure the temperature probe is properly stored in the
probe well.
2. Disconnect the temperature probe cable from the monitor.
3. Connect Temp simulator; set to 80.2°F.
4. Turn the monitor on.
5. If your measurement units are set to °C, switch the measurement units to °F.
“Temperature unit of measurement configuration settings”
page 3-10 for instructions.
6. Put the monitor into temperature monitor mode:
a. Remove the probe from the well.
b. Press the
+
key.
c. Verify that a blinking temperature appears quickly. If not, return the
probe to the holder and repeat steps 6a and 6b.
7. Record and verify the reading in the temperature display is 80.2°F ±0.2°F.
8. Set the simulator to 98.6°F.
9. Record and verify the reading in the temperature display is 98.6°F ±0.2°F.
10. Set the simulator to 107.8°F.
11. Record and verify the reading in the temperature display is 107.8°F ±0.2°F.
12. If necessary, switch the measurement units to °C. Refer to
of measurement configuration settings”
on page 3-10 for instructions.
13. Confirm all test results are recorded on the
14. Calibration is complete. If the monitor does not pass the calibration
verification, contact GE Technical Support.
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......