![GE CareScape V100 Скачать руководство пользователя страница 87](http://html.mh-extra.com/html/ge/carescape-v100/carescape-v100_service-manual_128562087.webp)
2048723-002C
CARESCAPE V100 Vital Signs Monitor
4-29
Maintenance:
Test results form
Test results form
Description
Min
Max
Actual
Pass-Fail-N/A
NIBP Tests
Pneumatic leakage result (mmHg)
0
6
Pressure transducer verification
Pressure reading at 200mmHg, top display -
Systolic
197
203
Pressure reading at 200mmHg, bottom display -
Diastolic
197
203
Pressure reading at 150mmHg, top display -
Systolic
147
153
Pressure reading at 150mmHg, bottom display -
Diastolic
147
153
Pressure reading at 100mmHg, top display -
Systolic
97
103
Pressure reading at 100mmHg, bottom display -
Diastolic
97
103
Pressure reading at 50mmHg, top display -
Systolic
47
53
Pressure reading at 50mmHg, bottom display -
Diastolic
47
53
Overpressure verification
Overpressure threshold, Adult (mmHg)
305
325
Overpressure threshold, Neonate (mmHg)
150
165
Buttons
NIBP Determination Initiated
‘
E80
’ displayed on
Systolic
display
Audible alarm can be silenced
Alarm silence indicator (bell) is lit
Overpressure alarm can be cleared
Alarm button is functioning
NIBP determination
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)
Heart rate reading (bpm)
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)
Heart rate reading (bpm)
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......