Technical specifications and default settings:
Specifications
2048723-002C
CARESCAPE V100 Vital Signs Monitor
A-9
D-YS & D-YSE
± 3.5 digits
D-YS & D-YSPD
± 3.5 digits
DS-100A
± 3 digits
OXI-A/N (adult)
± 3 digits
OXI-A/N (neonate)
± 4 digits
OXI-P/I
± 3 digits
Neonatal sensor accuracy
When sensors are used on neonatal subjects as
recommended, the specified accuracy range is increased
by ±1 digit, as compared to adult usage, to account for the
theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood. For example, MAX-N
accuracy on neonates is ±3 digits, rather than ±2 digits.
Sensor light source
Wavelength***
Infrared: 890 nm (nominal)
Red: 660 nm (nominal)
Power dissipation
Infrared: 22.5 mW (max)
Red: 30 mW (max)
*
The MAX-N, D-YS, OXI-A/N, and OxiCliq N were tested on patients >40 kg.
**
The accuracy specification has been determined between saturations of 80%-100%.
*** Information about wavelength range can be especially useful to clinicians.
Specifications: Nellcor sensor accuracy
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......