5-2
CARESCAPE V100 Vital Signs Monitor
2048723-002C
Troubleshooting:
Overview
Overview
The symptoms and solutions in this chapter represent only a few of the
problems that you may encounter and are not intended to cover every possible
problem that may occur.
A systematic approach to the diagnosis of problems as well as a general
understanding of the architecture, both hardware and software, of the monitor
are essential to ensure successful troubleshooting of a device. GE recommends
formal service training before repairs are attempted. These troubleshooting
procedures combined with training provide the service technician with skills
necessary to service and repair a monitor in the event of a malfunction.
Problems
Before starting any detailed troubleshooting, complete a thorough visual
inspection of the following.
All cable connections secure?
Devices properly powered?
Connected to a proper power source?
Problem
Possible reason
Solution
Exergen temperature does not
appear on the monitor
The host communication bit rate
may be set incorrectly.
Confirm that the host
communication bit rate (br) is set at
the default setting (
9600
). Refer to
page 3-11 on how to determine the
bit rate.
Loss of electrical contact between
the scanner and the monitor (e.g.,
corrosion on the connector inside the
scanner).
If the scanner appears to be working
properly, but will not communicate
with the monitor:
1. Verify that the connector is
properly connected.
2. Verify that the battery door is
securely fastened.
If there is still no communication with
the monitor:
1. Plug and unplug the connector
several times.
2. Rub off any corrosion on the
connector inside the scanner.
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......