3-4
CARESCAPE V100 Vital Signs Monitor
2048723-002C
Installation:
Powering the monitor
When about 5 minutes of main battery charge remains:
The low-priority
BATTERY LOW
alarm escalates to a high-priority
BATTERY LOW
alarm.
The
BATTERY LOW
indicator flashes.
Any NIBP determination in progress at the time of the alarm escalation is
allowed to finish.
Any
Stat
mode cycle that was initiated before the alarm escalation is
allowed to finish.
The user is not able to initiate:
any new NIBP determinations of any type
any printouts
NOTE
At this time, it is highly recommended to plug the monitor into external DC
power.
When 5 minutes of main battery charge expires:
After 5 minutes of high-priority
BATTERY LOW
alarm, the monitor enters a
battery low shutdown.
No error code is displayed.
The
BATTERY LOW
indicator flashes.
The monitor alarms for 2.5 minutes, then shuts down completely.
NOTE
You must plug the monitor into DC power before resuming monitoring.
After plugging the monitor into DC power:
The
BATTERY LOW
indicator (when the monitor is on) and
CHARGING
indicator illuminate.
The
BATTERY LOW
indicator turns off when the battery level reaches a
sufficient charge level to operate without the
BATTERY LOW
alarm active.
E13
BATTERY LOW
At any time while the high-priority
BATTERY LOW
alarm is active, certain actions
can trigger the ‘
E13
’
BATTERY LOW
alarm: any attempt to start an NIBP
determination or a printout. This alarm is giving you additional warning that the
battery charge is critically low.
NOTE
At this time, it is highly recommended to plug the monitor into external DC
power.
The ‘
E13
’
error code appears in the
min
window.
The
BATTERY LOW
indicator flashes.
This alarm can be acknowledged by pressing the
Silence
button.
The user is not able to initiate:
any new NIBP determinations of any type
any printouts
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......