4-16
CARESCAPE V100 Vital Signs Monitor
2048723-002C
Maintenance:
Calibration procedures and tests
Pressure transducer calibration
Perform only if Pressure Transducer Verification is out of tolerance as specified in
the
NOTE
To enter service mode press and hold
Cycle
button while pressing the
On/
Off
button.
1. Turn the monitor on and enter Service Mode.
2. The
min
window should display
0
.
3. Open valve on bulb to open pressure system to atmosphere.
4. Verify the manometer reads zero.
5. Press
Cycle
button and
1
should appear in the
min
window.
6. Close valve on bulb and slowly inflate pressure to 200 mmHg (using the
manometer as reference).
7. Press
Menu
button when pressure reads exactly 200 mmHg to set
calibration value.
8. To save the calibration setting, press the
Cycle
button until
6
appears in the
min
window.
9. Press and hold
Menu
button until monitor beeps and the number in the
MAP/Cuff
window decrements by 1, which acknowledges that data was
saved.
10. Turn the monitor off.
Overpressure verification
NOTE
To enter service mode press and hold
Cycle
button while pressing the
On/
Off
button.
1. Wait a few seconds after entering service mode. Press
Cycle
button so that
the
min
window changes from
0
to
1
.
2. Use the inflation bulb to inflate close to 300 mmHg. Slowly inflate (1 to 2
mmHg/sec) until valve opens and pressure is released.
3. Record and verify pressure at which valve opens on the
4. Press
Cycle
button and
2
should appear in the
min
window.
5. Use the inflation bulb to inflate close to 150 mmHg. Slowly inflate (1 to 2
mmHg/sec) until valve opens and pressure is released.
6. Record and verify pressure at which valve opens on the
7. Turn unit off.
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......