4-4
CARESCAPE V100 Vital Signs Monitor
2048723-002C
Maintenance:
Cleaning
Cleaning the exterior surfaces of the monitor, monitor
accessories, or the Exergen temporal scanner
Disconnect the monitor from AC power before cleaning or disinfecting its
surface. The exterior surfaces of the monitor, monitor accessories, and temporal
scanner may be cleaned with a dampened, lint-free cloth. Wipe off all cleaning
solutions with a clean, dry cloth and let air dry for at least 15 minutes. Use one of
the following approved solutions:
Water
Mild soap
Household bleach (5.25% sodium hypochlorite). Mix 10:1 with distilled water.
Sagrotan (dilution 3:100, containing 75 mg tartaric acid per 100 ml solution).
Exergen temporal scanner only: Alcohol-based cleaning agents can be used
on the scanner’s probe head and metal neck only.
Never use the following cleaning agents on the monitor or the Exergen temporal
scanner:
Abrasive cleaners or solvents of any kind
Acetone
Ketone
Betadine
Alcohol-based cleaning agents. (However, an alcohol-based cleaning agent
can be used on the Exergen scanner’s probe head and metal neck only.)
Petroleum-based cleaning agents
Any type of solution that contains ammonium chloride, conductive
solutions, wax or wax compounds
Sodium salts
NOTE
Never autoclave or steam clean the monitor, cuffs, or accessories.
Cleaning the displays of the monitor or Exergen temporal
scanner
To clean the display covers, use a soft, clean cloth dampened with a glass
cleaner. Never spray the glass cleaner directly onto the display, and never use
alcohol- or petroleum-based products.
Cleaning the sensor lens of the Exergen temporal scanner
Dirt, greasy film, or moisture on the scanner lens will interfere with the accuracy
of the temperature reading. Regularly clean the lens with a cotton swab dipped
in alcohol and follow the instruction label on the scanner. Only use gentle
pressure for cleaning to avoid lens damage. Water can be used to remove any
residual film left by the alcohol. Do not use bleach or other cleaning solutions on
the sensor lens.
Cleaning the probe head and neck of the Exergen temporal
scanner
Use an alcohol-based cleaning agent on the Exergen scanner’s probe head and
metal neck only.
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......