Installation:
Configuring the monitor
You can then press the
Menu
button to cycle through all the advanced
configuration mode option pages. After all options pages have been displayed,
the display will return to the first advanced configuration mode page (displaying
ACF
). You can use the
+
and
-
buttons to make changes to settings. After making
changes, simply cycle the power to return to normal operation mode. Changes
are automatically be retained.
The advanced configuration mode option pages are as follows:
Service mode
Service mode is entered by holding the
Cycle
button while powering on with the
On/Off
button. You can press the
Cycle
button to advance through the available
service mode pages. You can use the
+
and
-
buttons to make changes to
settings.
NOTE
Only transducer calibration pages are available until calibration is valid.
Save settings
Calibration and other service mode setting changes will not be retained unless
the “Save Settings” operation is executed (on the final Service Mode options
page). To save settings in service mode, advance to the last page (page 6), then
press and hold the
Menu
button until the second of two tones sound.
Displayed on monitor
Function
ACF
Advanced Configuration Mode announcement
(No settings are entered on this page.)
rEn
Remote mode
0
: Remote mode is disabled
1
: Remote mode is enabled (default)
Adr
Host communication unit address
32
: This is the default value: 126 max.
br
Host communication bit rate (bits per sec)
0
: 300 bps
1
: 600 bps
2
: 1200 bps
3
: 2400 bps
4
: 4800 bps
5
: 9600 bps (default)
6
: 19200 bps
NOTE
For the Exergen temperature to appear on
the monitor, the host communication bit rate
must be set to
9600
bps.
nod
Host communication mode
0
: Host communication command mode (default)
1
: 1846 Compatibility mode
(1846 mode requires user to also select 600 bps.)
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......