2048723-002C
CARESCAPE V100 Vital Signs Monitor
5-3
Troubleshooting:
Alarm code interpretation
Alarm code interpretation
General system error codes are listed in this section. If any other system or
similar code appears, record the error message and report the failure to GE
Technical Support. Refer to the operator’s manual for information about patient
alarms and general procedural alarms.
System failures
System failure mode is entered when the monitor has a depleted battery, or a
hardware or software failure. A distinctive alarm tone is generated for up to 5
minutes, after which if the monitor isn’t turned off, it shuts down completely.
When a system failure is encountered, the error code is displayed on the screen
for > 5 seconds and the system enters FAILSAFE mode. The error code is logged
in the history log.
If the
On/Off
button remains depressed for between 10 and 20 seconds, a
special situation occurs. It forces a reset and produces a fatal alarm without a
displayed code.
Alarm conditions and error codes
When responding to a monitor alarm, always
check the patient first
and then
check the monitor, cuff, hose and sensors. Press
Silence
to clear patient alarm
conditions.
Error log
You can view and print an error log that stores up to 40 error code entries. The
log is a “rolling” list that—once 40 entries are stored—deletes the oldest entry in
order to add the most recent entry. The error log is saved until the monitor
experiences a memory loss, then all entries are deleted.
Procedure to view and print error code history log
1. Enter the Advanced Configuration Mode (ACF) by holding down the
Minus
(
-
) and
Menu
buttons while powering up the monitor (pressing the
On/Off
button).
2. The monitor briefly displays the software revision, then displays
ACF
in the
Systolic
window.
3. The monitor is now in Advanced Configuration Mode.
4. To view the error log use the
History
button to step through the log in
reverse order of when the error occurred (oldest appears first).
5. The
Systolic
window shows the Year the error occurred.
6. The
Diastolic
window shows the Day the error occurred.
7. The
MAP
window shows the Month the error occurred.
8. The
min
window shows the Time the error occurred.
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......