CARESCAPE V100 Vital Signs Monitor
4-17
Maintenance:
Calibration procedures and tests
Button testing
1. Disconnect the cuff/hose assembly and power on the unit.
2. Press
Inflate/Stop
button.
3. Verify a NIBP determination has been initiated.
4. Block pump port and verify
‘
E80
’
alarm.
5. Press
Silence
button, verify alarm has been cleared.
6. Verify that the alarm silence indicator (bell) is lit.
7. Press
Alarm
button several times, verify unit cycles through all alarm
settings (i.e.,
SYS
,
DIA
,
SpO2
).
8. Confirm all test results are recorded on the
9. Turn unit off.
NIBP functional test
NOTES
The NIBP function test does not test for measurement accuracy.
While an NIBP simulator device may be useful to verify that the monitor
responds to oscillometric pulsations, it should not be used as a basis for
assessing the accuracy of measurement. For more information, see
“Appropriate use of NIBP simulators”
1. Remove the calibration setup and attach an adult cuff and hose to a
simulator (be sure to select the correct cuff size).
2. Press
Inflate/Stop
button on monitor to begin a determination.
3. Record
Systolic
,
Diastolic
,
MAP
and heart rate from the monitor display.
4. Wait 1 minute, then press
Cycle
button to initiate a determination in Auto
NIBP mode.
5. Record
Systolic
,
Diastolic
,
MAP
and heart rate from the monitor display.
6. Wait 1 minute, then press
Cycle
button until
stat
is displayed in the
min
window to initiate a determination in
STAT
mode.
7. Record
Systolic
,
Diastolic
,
MAP
and heart rate from the monitor display.
8. Press
Inflate/Stop
button, end
STAT
mode.
9. Confirm all test results are recorded on the
NIBP overpressure verification
1. Remove the cuff/hose from the monitor, restrict airflow from cuff hose port.
2. Press
Inflate/Stop
to begin NIBP determination.
3. Verify
‘
E80
’
is displayed on the
Systolic
window and audible alarm sounds.
4. Remove the air restriction.
5. Press
Inflate/Stop
and verify that the pump does not start.
6. Press the
Silence
button.
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......