Maintenance:
Visual inspection
Visual inspection
The monitor and its components should be carefully inspected prior to
installation, once every 12 months thereafter and each time the equipment is
serviced.
WARNING
Do not use damaged sensors, cables, or connectors.
Carefully inspect the equipment for physical damage to the case, the display
screen, and the keypad. Do not use the monitor if damage is determined.
Refer damaged equipment to qualified service personnel.
Inspect all external connections for loose connectors or frayed cables.
Have any damaged connectors or cables replaced by qualified service
personnel.
Inspect the display face for marks, scratches, or other damage.
Safety labels and inscription on the device are clearly legible.
Cleaning
Cleaning schedule
CAUTION
To prevent cross-contamination, clean the exterior surfaces of
the monitor, monitor accessories, and reusable sensors on a
regular basis in compliance with your institution’s infection
control unit and/or biomedical department’s local policy.
Cleaning the monitor, monitor accessories, and the Exergen temporal scanner
WARNING
Never pour or spray water or any cleaning solution on the
equipment or permit fluids to run behind switches, into
connectors, into the recorder, or into any ventilation openings
in the equipment. Do not let fluid “pool” around connection
pins.
WARNING
Use of unapproved cleaning agents can cause case damage
resulting in unintended fluid ingress and a potential for
compromising electrical safety.
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......