A-2
CARESCAPE V100 Vital Signs Monitor
2048723-002C
Technical specifications and default settings:
Specifications
Specifications
General
General specifications
Mechanical
Dimensions
Height
19.5 cm (7.7 in)
Width
21.9 cm (8.6 in) without temperature
25.4 cm (10.0 in) with temperature
Depth
13.5 cm (5.3 in)
Weight (Including battery)
2.4 kg (5.4 lb)
Mountings
Self-supporting on rubber feet, pole mounted, or wall mount bracket
Portability
Carried by recessed handle
Power requirements
Power converter universal
P/N: 2018859-001
Protection against electrical shock
Class II
AC input
100 to 240VAC, 0.5A
DC output voltage
12VDC at 1.0A
The AC mains power adapter contains a nonresettable and
nonreplaceable fuse.
Monitor
Protection against electrical shock
Internally powered or Class II when powered from specified external
power supply.
DC input voltage
12 VDC, supplied from a source conforming to IEC 60601-1.
Fuses
The monitor contains three fuses. The fuses are mounted within the
monitor. The fuses protect the low voltage DC input, the main battery,
and the remote alarm output. The +5 V output on the host port connector
is regulated by internal supply.
Main Battery
Refer to
Environmental
Operating temperature
+ 5°C to + 40°C (+ 41°F to + 104°F)
Operating atmospheric pressure
500 hPa to 1060 hPa
Содержание CareScape V100
Страница 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Страница 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Страница 42: ...2 18 CARESCAPE V100 Vital Signs Monitor 2048723 002C Equipment overview Functional description For your notes ...
Страница 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Страница 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Страница 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Страница 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Страница 98: ...5 8 CARESCAPE V100 Vital Signs Monitor 2048723 002C Troubleshooting Alarm code interpretation For your notes ...
Страница 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Страница 127: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 29 Parts lists and drawings Field replaceable units FRUs ...
Страница 128: ...6 30 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 129: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 31 Parts lists and drawings Field replaceable units FRUs ...
Страница 130: ...6 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 131: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 33 Parts lists and drawings Field replaceable units FRUs ...
Страница 132: ...6 34 CARESCAPE V100 Vital Signs Monitor 2048723 002C Parts lists and drawings Field replaceable units FRUs ...
Страница 133: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 35 Parts lists and drawings Field replaceable units FRUs ...
Страница 145: ...2048723 002C CARESCAPE V100 Vital Signs Monitor A 1 A Technical specifications and default settings ...
Страница 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Страница 177: ......