Optional Display and Pushbuttons
The optional display and buttons (Figure 6) are mounted on the SVI FF cover plate. The three
pushbutton switches operating in conjunction with the display permit reading and
modification of the instrument operating parameters without a PC or hand-held
communicator. These switches perform generic functions - Increase, Decrease, and Accept by
movement through a conventional menu structure, see“Using the Pushbuttons and Digital
Interfaces” on page 127. The switches are operated in a hazardous environment without
compromising the flameproof enclosure.
Figure 6 Optional Display
Link Active Scheduler
To use the SVI FF in a link active scheduler configuration refer to the documentation for the
specific DCS.