Revision 14
67
August 02, 2019
•
MODBUS ADD
- Presents the current Modbus unit address. This parameter can be edited
by pressing the EDIT key, typing in a new value followed by the ENTER key.
•
CURRENT
- Presents the sensor LED current in mA.
•
TEMP
- indicates current temperature of the sensor block.
•
SENSOR
- Presents the sensor firmware revision number and cannot be edited.
•
FIRM
- indicates the analyzer firmware revision number and cannot be edited.
•
DIAGNOSTICS
- If the
Diagnostics
option is selected, a series of hardware diagnostics will
be performed when the analyzer is powered up. If Diagnostics is not selected, the hardware
diagnostics will not be performed. This function (ON/OFF) can be edited by pressing the
EDIT key followed by the ENTER key.
•
COLD_R
- resets all configuration parameters to the factory default. This will take about 1
minute and the analyzer will reboot. This function (ON/OFF) can be edited by pressing the
EDIT key followed by the ENTER key.
•
SEN. CAL
- This field is used to initiate a sensor calibration (see Section 7.3).
4.8.2 The DIO Menu (Digital Inputs/Outputs)
When DIO (F2) is pressed on the
Configuration
page (Figure 4-12), the
DIO
display which
describes the Digital Outputs appears as shown in Figure 4-14.
mV Run State Time
6 1H1 S1 115
FURN OFF DI-POL Edit
PRESS OFF DI-POL Edit
CAL OFF DI-POL Edit
WET 1 OFF DI-POL
Edit
F1 F2 F3 F4
PGUP Up Down PGDN
Figure 4-14: The DIO Display
The status of the items on this menu can be edited by pressing the EDIT key followed by the
ENTER key. Each item can be edited (e.g. there are 4 wet lines 4 dry lines, etc); and can be
accessed via the function keys). The items on this menu are listed in Table 4-6.
Table 4-6: DIO Menu
FURNACE
DI-POL
(a)
PRESSURE
DI-POL
(a)
CALIBRATION
DI-POL
(a)
WET 1-4
DI-POL
(a)
DRY 1-4
DI-POL
(a)
RELAY 1-6
FS-POS
(b)
RELAY 1-6
LCH_EN
(c)
(a) DI-POL indicates the polarity of each of the available discreet inputs. See Section 5.11 and 5.12.
(b) FS-POS shows the fail safe position of each of the six relays.
(c) LCH_EN indicates if the alarm associated with each relay is latching or non-latching.
Содержание ProTech903
Страница 2: ......
Страница 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Страница 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Страница 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Страница 167: ...Revision 14 167 August 02 2019 Figure 9 1 DC Power Wiring Diagram Figure 9 2 AC Power Wiring Diagram...
Страница 168: ...Revision 14 168 August 02 2019 Figure 9 3 Total Sulfur Wiring Diagram Figure 9 4 AC DC Power Wiring Diagram...
Страница 169: ...Revision 14 169 August 02 2019 Figure 9 5 Solenoid Drivers Wiring Diagram Figure 9 6 Relays Wiring Diagram...
Страница 173: ...Revision 14 173 August 02 2019 Figure 9 13 Isolated RS 485 Port Figure 9 14 P17 RS232 Port to Ethernet Cable...
Страница 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Страница 190: ...Revision 14 190 August 02 2019...