Revision 14
51
August 02, 2019
Figure 3-22: Minihelic Indicating a Good Seal (L); Connected to Vent Elbow (R)
d) Check the pressure indicated on the gauge. If the pressure reads 4 or above, as shown
in Figure 3-22, the tape is well sealed against the sample chamber by the compression
head. Press
Tape Advance
on the keypad and ensure that the pressure does not drop
below 3 even when the tape is moving. It should return to 4 by the time the tape stops
moving.
If the pressure is NOT 4 or greater, there is a problem with the seal somewhere in the
chassis.
3.7.3 Leak Testing the Total Sulfur Accessory
Connect the USB to USB cable from the laptop to the analyzer and log onto the
analyzer using the Galvanic GUI software. Go to the Global tab and turn off the power
to the furnace by setting the
Furnace Pulse Width
to 0 %. Perform a temporary write to
the analyzer and wait until the furnace housing is cool to the touch.
test the furnace for leaks as described below:
a) Turn the hydrogen flow on. Ensure hydrogen is flowing at 2.0 and sample is flowing at 1.0
on the rotameters.
b) Place a ¼ inch cap on the top of the furnace output flame arrestor as shown in Figure 3-
24. This will block flow through the total sulfur furnace.
Содержание ProTech903
Страница 2: ......
Страница 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Страница 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Страница 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Страница 167: ...Revision 14 167 August 02 2019 Figure 9 1 DC Power Wiring Diagram Figure 9 2 AC Power Wiring Diagram...
Страница 168: ...Revision 14 168 August 02 2019 Figure 9 3 Total Sulfur Wiring Diagram Figure 9 4 AC DC Power Wiring Diagram...
Страница 169: ...Revision 14 169 August 02 2019 Figure 9 5 Solenoid Drivers Wiring Diagram Figure 9 6 Relays Wiring Diagram...
Страница 173: ...Revision 14 173 August 02 2019 Figure 9 13 Isolated RS 485 Port Figure 9 14 P17 RS232 Port to Ethernet Cable...
Страница 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Страница 190: ...Revision 14 190 August 02 2019...