Revision 14
143
August 02, 2019
6.5 Calibrating/Validating the Analyzer Outputs
The
Maintenance
tab (Figure 6-11) is used to manually manipulate the analyzer’s outputs.
The Analog output loop test can be performed by using the “Analog Output Test” feature to
deliver 0, 25, 50, 75 and 100% of full scale or any other test values. The outputs can also be
adjusted by using the
Analog Output X Calibration
feature. A complete discussion of the use
of this tab is given in Section 5.19.
Figure 6-11: Maintenance Tab