11.3 Communications / Interface
Outputs
6 - 4-20mA outputs (isolated and scaled to
range)
3 - SPDT alarm relays 5 amps @ 30 VDC
or 8amps @ 120/230 VAC (up to 3 more
optional)
1 - 3 amp max 35 VDC or 3A 24-280 VAC
Solenoid Driver (standard), up to 7 more
(optional)
1 - USB port
1 - RS-232 Serial Port
1 - RS-485 isolated serial Port
1 - LAN Port for extended analog outputs
1 - Ethernet (optional)
Modbus communication on all ports
Inputs
8 discrete inputs (4 - Dry contact digital inputs
& 4 - wet contact digital inputs (12/24
VDC)
2 - 4-20 mA, user selected
Operator Interface
128x64 LCD Screen
Handheld Keypad (Intrinsically safe keypad
for D1 models)
10 visual indicators (LED) for quick status
Remote PC application program
Содержание ProTech903
Страница 2: ......
Страница 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Страница 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Страница 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Страница 167: ...Revision 14 167 August 02 2019 Figure 9 1 DC Power Wiring Diagram Figure 9 2 AC Power Wiring Diagram...
Страница 168: ...Revision 14 168 August 02 2019 Figure 9 3 Total Sulfur Wiring Diagram Figure 9 4 AC DC Power Wiring Diagram...
Страница 169: ...Revision 14 169 August 02 2019 Figure 9 5 Solenoid Drivers Wiring Diagram Figure 9 6 Relays Wiring Diagram...
Страница 173: ...Revision 14 173 August 02 2019 Figure 9 13 Isolated RS 485 Port Figure 9 14 P17 RS232 Port to Ethernet Cable...
Страница 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Страница 190: ...Revision 14 190 August 02 2019...