Revision 14
146
August 02, 2019
h) When the calibration is complete, turn the 3 way valve back to process gas.
i) Press the
Bypass
button on the keypad to take the analyzer out of Bypass mode so that
any alarm outputs will be re-activated.
6.6.3 Auto Calibration
If this option is employed, the analyzer is equipped with a 3 way solenoid valve that is
controlled by the analyzer and is used to introduce calibration gas to the analyzer.
To perform an auto calibration:
a) Initiate an Auto Calibration run, either from the GUI or from the keypad. The Auto
Calibration can be initiated from the GUI by pressing the
Stream 1 Calibration
button as
shown in Figure 6-12. The Auto Calibration can be initiated from the keypad by selecting
the stream to be calibrated, and scrolling to CAL. The CAL entry will indicate that the
status is OFF. Press the EDIT key on the keypad and press the ENTER key to toggle the
value to ON.
Note that when an Auto Calibration is initiated, either from the GUI or from
the keypad, the analyzer will finish the current analysis cycle before it
begins the calibration cycle.
Figure 6-12: Calibration Tab
b) The analyzer will run the Auto Calibration according to the setup of the
Stream/Calibration run. The analyzer will run the total number of runs defined by
Содержание ProTech903
Страница 2: ......
Страница 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Страница 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Страница 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Страница 167: ...Revision 14 167 August 02 2019 Figure 9 1 DC Power Wiring Diagram Figure 9 2 AC Power Wiring Diagram...
Страница 168: ...Revision 14 168 August 02 2019 Figure 9 3 Total Sulfur Wiring Diagram Figure 9 4 AC DC Power Wiring Diagram...
Страница 169: ...Revision 14 169 August 02 2019 Figure 9 5 Solenoid Drivers Wiring Diagram Figure 9 6 Relays Wiring Diagram...
Страница 173: ...Revision 14 173 August 02 2019 Figure 9 13 Isolated RS 485 Port Figure 9 14 P17 RS232 Port to Ethernet Cable...
Страница 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Страница 190: ...Revision 14 190 August 02 2019...