5.17.2 Analog Input Parameters and Controls
•
Description
– used to enter the Analog Input descriptor. This information will appear under
the Analog Inputs section on the Status Tab and on the analyzer LCD display
•
Minimum
– the lower limit of the range (typically zero).
•
Range
– the maximum or span limit of the range. If the value in the
Minimum
box is not
zero, the Minimum and Range values are added together to give the full scale value.
•
Value
– the calculated range value for a given 4 – 20 mA input.
•
Engineering Units
– the units that correspond to the type of analog input (e.g. ppm, %,
psig, etc).
•
Calibrate A. I. #
– when this button is pressed the
Scaled Zero
and
Scaled Span
fields will
be activated and the appropriate values can be entered.
•
Capture Zero
– used to capture the A/D bits that are associated with a 4mA input during
the calibration of the analog input.
•
Scaled Zero
– used to enter the zero value or the range. For a 4 – 20 ma range, the 4 mA
value would be entered.
•
Raw Zero
– shows the A/D bit count for the 4mA input
•
Capture Span
– used to capture the A/D bits that are associated with a 20mA input during
the calibration of the analog input.
•
Scaled Span
– used to enter the span value or the range. For a 4 – 20 ma range, the 20
mA value would be entered.
•
Raw Span
– shows the A/D bit count for the 20 mA input
5.17.3 Analog Inputs Alarm
The
Analog Inputs Alarm
fields are used to select the input alarms to be used, the limit for the
input and the solenoid or relay to be used for that alarm. The appropriate check box should
be selected for each input.
Содержание ProTech903
Страница 2: ......
Страница 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Страница 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Страница 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Страница 167: ...Revision 14 167 August 02 2019 Figure 9 1 DC Power Wiring Diagram Figure 9 2 AC Power Wiring Diagram...
Страница 168: ...Revision 14 168 August 02 2019 Figure 9 3 Total Sulfur Wiring Diagram Figure 9 4 AC DC Power Wiring Diagram...
Страница 169: ...Revision 14 169 August 02 2019 Figure 9 5 Solenoid Drivers Wiring Diagram Figure 9 6 Relays Wiring Diagram...
Страница 173: ...Revision 14 173 August 02 2019 Figure 9 13 Isolated RS 485 Port Figure 9 14 P17 RS232 Port to Ethernet Cable...
Страница 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Страница 190: ...Revision 14 190 August 02 2019...