Revision 4
138
April 2020
Section 10
Specifications
Performance Specifications
Auto Calibration
User Selectable Frequency
Zero Drift
< 0.25% of full scale per day (based on
autozero frequency of once/hour
Accuracy
+/- 1% of Full Scale with pressure
compensation (with all species present)
Sensitivity
+/- 1.0 % of Full Scale
Repeatability
Better than 0.5% of full scale (with all species
present)
Response Time
Analyzer
– Near Instantaneous
Total System
– typically less than 30 seconds
(depends on probe length)
Physical Specifications
Size - Outside Dimensions
(including Mounting Frame)
57” W x 41.25” H x16” D
(145 cm x 105 cm x 41 cm)
Analyzer cabinets
Two cabinets, each 24” W x 30” H x 12” D
(61 cm x 76.2 cm x 30 cm)
Weight (Total System on Frame
Approximately 275 lbs (125 kg),varies with
options
Services Required
Electrical
100-240 Vac, 1 phase, 50/60 Hz, 800 W
Instrument Air
17 SCFM at 80 psi (maximum case with Vortec
Cooler)
Steam
Nozzle
– 50 psi
Area Classification
All specifications are subject to change as part of our ongoing product
improvement program.
943-TGX-CE (ATEX)
Device Group:
II 2 G.
Ex protection:
II 2 G Ex d e ia mb px IIC T3 Gb
IP Protection:
IP65
Ambient Temperature:
-20°C to +60°C.
943-TGX-CE (IECEx)
Device Group:
II 2 G.
Ex protection:
II 2 G Ex db eb ia mb pxb IIC T3 Gb
IP Protection:
IP65
Ambient Temperature:
-20°C to +60°C.
Содержание 943-TGX-CE
Страница 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Страница 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Страница 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Страница 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Страница 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Страница 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...