Revision 4
10
April 2020
Important Safety Guidelines for the 943-TGX-CE
Field Programmable Tail Gas Analyzer System
Please read the following warnings and cautions carefully before using the
943-TGX Tail Gas Analyzer
System:
This equipment must be used as specified by the manufacturer or overall safety will
be impaired.
Access to this equipment should be limited to authorized, trained personnel ONLY.
Due to the thermal mass of the hardware, cooling of the items takes substantial
time.
Observe all warning labels on the analyzer enclosures.
Install fuses of the Type and Rating as shown on the Fuse Identifier labels.
The analog outputs and alarm relay contacts may be powered by a source separate from the
one (s) used to power the analyzer system. Disconnecting the AC Mains Source (s) may not remove
power from the analog output signals.
Any safety recommendations or comments contained herein are suggested guidelines only.
Galvanic Applied Sciences Inc. bears no responsibility and assumes no liability for the use and/or
implementation of these suggested procedures.
This system, when operating in its normal mode, and/or when it is being serviced, maintained,
installed and commissioned contains items which may be hazardous to humans if handled or
operated incorrectly or negligently. These items include, but are not limited to:
High Voltage Electrical Energy
Toxic and Explosive Gases
Intense Ultraviolet Radiation
High Temperature Surfaces
Installation of the system requires the opening of the process sample point to allow for the insertion
of the system sample probe assembly. To achieve this, removing the process access point blind
flange is necessary. When the flange is removed, toxic, hot (approximately 150ºC/300ºF) gases
and molten sulfur may be expelled to the atmosphere until the system sample probe and its
associated flange are in place and securely fastened.
It is recommended that the personnel installing the probe wear plant approved breathing air
apparatus, and approved personal protective equipment (i.e. gloves, coveralls and protective eye
Содержание 943-TGX-CE
Страница 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Страница 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Страница 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Страница 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Страница 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Страница 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...