Revision 4
101
April 2020
2. Place the analyzer in
Offline
mode using the Online / Offline control on the
Analysis
2
panel. (Refer to Section 4.3.1)
3. Place the analyzer in back-purge by clicking the
Back Purge
check box on the
Analysis 1
panel OR by pressing the F6 key on the keypad. The displayed
concentration and air demand outputs should stabilize to a near zero readings within
seconds. Use a stopwatch to measure the time required for the readings to drop from
stable process values to stable zero values. Record the stabilized zero concentration
output values.
The offset from zero will vary based on length of time since the last zero
calibration cycle (manually initiated or automatic).
4. On the
Spectrum
panel, observe the raw scan signal and compare the shape of the
scan trace to the
Factory Reference
. Note any extreme deviations from the factory
referene. Record the value in the
Peak Height
field in approximate A/D counts
(±1,000).
5. If the value in the
Peak Height
field is less than 16,000, the signal peak height should
be optimized using the
Optimize Integ. Time
button. Refer to Section 4.6.
6. Return to the
Analysis 1
panel. Ensure that the
Status
indicator is green (Normal) and
the Service indicator is also green (Normal). Check the displayed concentration values.
If they are not very close to zero (±0.02% for all indicated values) a
Manual Zero
should be performed. Refer to Section 4.3.4.
7. Return the analyzer to
Sample
mode by clicking on the
Sample
checkbox on the
Analysis 1
panel OR by pressing the F6 key on the keypad. Observe and record how
long it takes for the concentration readings to stabilize at the process value. Record
the stabilized process values. A long response time to a stabilized sample reading may
indicate plugging of the sample system or filter, or inadequate aspirator drive air flow.
Record any observed anomalies and/or valve position adjustments made.
The response time from back purge to stabilized sample values should be
>30 seconds, but the exact value depends on a variety of factors, including
probe length and instrument air pressure.
8. Return the analyzer to
Online
by toggling the
Online / Offline
control back to the green
Online
state.
Содержание 943-TGX-CE
Страница 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Страница 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Страница 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Страница 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Страница 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Страница 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...