Revision 4
87
April 2020
Many of the parameters on this page are also available for editing
on the analyzer’s local
display using the analyzer keypad, though not all of them are. Table 24 gives an overview of
the parameters available for editing on this page.
Table 24: Configuration Parameters
Parameter
Editable
via
Keypad
Local Display
Editing Location
Explanation
Cell Temperature Set
Point
Y
Config
– Timers/Alarms
Control setpoint for analyzer’s measurement
cell heaters.
Low Cell Temperature
Alarm
Y
Config
– Timers/Alarms
Cell temperature below which a cell
temperature fault will be triggered
High Cell Temperature
Alarm
N
N/A
Cell temperature above which a cell
temperature fault will be triggered
Cell Temperature
Proportional Band
Y
Config
– Timers/Alarms
Cell temperature PID control parameter
Cell Temperature Reset
Time
N
N/A
Cell temperature PID control parameter
Probe Temperature Set
Point
Y
Config
– Timers/Alarms
Control setpoint for probe cooler solenoid
actuation
NOTE: Only if probe control is set to RTD
Probe Temperature
Dead Band
Y
Config
– Timers/Alarms
Number
of
degrees
below
probe
temperature setpoint measured probe gas
temperature must drop prior to probe cooler
solenoid turning off
NOTE: Only if probe control is set to RTD
Low Probe Temperature
Alarm
Y
Config
– Timers/Alarms
Probe gas temperature below which low
probe temperature fault will be triggered
High Probe Temperature
Alarm
Y
Config
– Timers/Alarms
Probe gas temperature above which high
probe temperature fault will be triggered
Cabinet Temperature Set
Point
Y
Config
– Timers/Alarms
Control cabinet temperature above which
the cabinet cooler will be turned on
Cabinet Temperature
Dead Band
Y
Config
– Timers/Alarms
Number
of
degrees
below
cabinet
temperature
setpoint
the
cabinet
temperature must drop prior to the cabinet
cooler switching off
Low Cabinet
Temperature Alarm
N
N/A
Control cabinet temperature below which a
cabinet
temperature
warning
will
be
triggered
High Cabinet
Temperature Alarm
N
N/A
Control cabinet temperature above which a
cabinet
temperature
warning
will
be
triggered.
High SVAP Warning
Y
Config
– Timers/Alarms
Sulfur vapour concentration above which a
High Sulfur Vapour warning will be triggered
High SVAP Alarm
Y
Config
– Timers/Alarms
Sulfur vapour concentration above which a
high sulfur vapour fault will be triggered
SVAP Set Point
N
N/A
Sulfur vapour concentration above which the
probe cooler will be turned on
NOTE: Only if probe control is set to
SVAP
SVAP Dead Band
N
N/A
Amount the sulfur vapour concentration
must drop below the set point before the
probe cooler switches off
NOTE: Only if probe control is set to
SVAP
Filter Points
N
N/A
Number of data points averaged together to
produce the displayed concentration values
Auto Cal Intervall (min)
Y
Config
– Timers/Alarms
Time interval from last zero calibration cycle
(manual or automatic) to next scheduled
auto zero calibration cycle.
Содержание 943-TGX-CE
Страница 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Страница 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Страница 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Страница 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Страница 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Страница 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...