Revision 4
32
April 2020
Section 3
Operation
3.1
Overview
The 943-TGX analyzer system is designed to operate automatically with a minimum need for
human intervention. Once installation, commissioning and initial start-up have been completed,
the only human intervention required is when the system indicates a problem or when routine
preventative maintenance is scheduled. Section 4 of this manual describes the local user
interface (UI) displayed on the analyzer’s full-colour display and its operation using the handheld
keypad.
The analyzer system computer automatically performs all the normal operational procedures
including sample flow initiation, sampling and analysis,
back purge
and
zero calibration
, range
sensitivity selection,
fault
detection, temperature zone control, and fail-safe back purge in the
event of an analyzer fault. Operators are alerted to an abnormal state or fault occurrence
through indicators on the local display located on the control cabinet door as well as fault and
warning
relay
contact closures.
The only manual adjustments to system controls (not through the computer) are the flow control
of various instrument air streams. Flow adjustment valves for the zero/purge gas flow rate, the
aspirator drive air flow rate and the probe cooler (used for sulfur vapour condensation at the
probe tip) air flow rate are located in the oven cabinet, as shown in Figure 12.
Figure 12: Location of Air Flow Control Valves in Oven Cabinet
Содержание 943-TGX-CE
Страница 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Страница 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Страница 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Страница 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Страница 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Страница 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...