10.11 Monitor Offset Calibration
The monitor option provides up to 15 additional process inputs, which can be configured from any combination
of thermocouple, analog and RTD inputs from the optional monitor input modules. The monitor input option
can also be configured to display the second input from compatible loop controllers attached to the system (one
from each available control loop).
The calibration range is fixed within the thermocouple, analog and RTD monitor input cards; however, the user
can input a linear offset in order to adjust for sensor tolerances and lead affects. When the second input of a
control loop is used for the monitor input, the loop controller provides calibration, range and input bias
adjustments. The input bias of the loop control can be used to set the linear offset for the input, or the monitor
offset calibration of the MCT can be used.
The left and right navigation buttons at the upper right of the screen allow the user to scroll through each
available monitor input. To adjust the input reading, enter the offset value in the ‘Offset’ field and verify the
reading. The offset range is -32760 to 32760 for a decimal point resolution of zero, -3276.0 to 3276.0 for a
decimal point of one, -327.60 to 327.60 for a decimal point of two and -32.760 to 32.760 for a decimal point
resolution of three. Be sure to press the ‘Save’ button to save the new offset value before proceeding onto
other inputs or exiting the monitor input offset screen or the offset value will revert back to its original value.
NOTE:
If using the monitor input offset of the MCT to offset the reading for the second input of a loop control,
the value for the monitor input as shown on the MCT
will not match that on the loop controller’s display.
The monitor input offset of the MCT only offsets the reading on the display, not on the loop control.
Therefore, to avoid confusion when using the second input of the loop controls for monitor points, it
is recommended that any input offset is applied through the loop controller’s input bias setting and
not the monitor input offset of the MCT.
10.12 Display Settings
The Display screen provides access to the touch screen calibration utility, back light settings and alarm volume
adjustment.
Содержание MCT-CM
Страница 1: ...MCT CM User Manual MCT CM User Manual Rev C August 2022 V1 2eCM30 V1 3eCM22 Supersedes May 2021...
Страница 20: ...MCT CM 20...
Страница 24: ...MCT CM 24...
Страница 45: ...MCT CM 45...
Страница 187: ...MCT CM 187...
Страница 227: ...MCT CM 227 A 4 Control Module Specifications Main CPU...
Страница 228: ...MCT CM 228...
Страница 229: ...MCT CM 229...
Страница 230: ...MCT CM 230...
Страница 231: ...MCT CM 231 Analog I O Card...
Страница 232: ...MCT CM 232...
Страница 233: ...MCT CM 233 Digital Input Card...
Страница 234: ...MCT CM 234...
Страница 235: ...MCT CM 235 Digital Output Card...
Страница 236: ...MCT CM 236...
Страница 237: ...MCT CM 237 Dual Communications Card Barcode and Remote Display Communication Cartridges...
Страница 238: ...MCT CM 238 A 5 Power Supply Specifications...
Страница 239: ...MCT CM 239...
Страница 240: ...MCT CM 240 A 6 Power On Delay Relay Specifications GE1A C10HA110...
Страница 241: ...MCT CM 241...