NOTE:
The set point entry is critical in order for the step to operate correctly when Edge is selected as the
wait type. During the wait state, each selected loop and/or monitor input must arrive at or cross the
wait set point in order for the program to resume operation. Thus, the previous step must ensure that
the process values will be moving towards the target wait set point prior to the wait step beginning. If
all selected loops and/or monitors do not cross the wait set point after the wait step has started, the
program will pause indefinitely on the step until the user takes action.
The digital input selections are
made by pressing ‘Digital’ button on the Wait screen. When using the digital
input wait for selection, the digital input must be configured as a
‘wait input’. If it is not configured as a ‘wait
input
’, the program will wait on the step indefinitely until the user takes action since the input is unable to trigger
the end of the wait condition. See section 10.7 Digital Inputs for information on setting the digital input function.
6.1.3.1
Delta Control Setting
The delta control setting, accessed by pressing the
“Delta” button, is a special feature of MCT that works in
conjunction with the loop and monitor wait for setting. Typical uses for this function include wood drying or
heat-treating processes (metal, food, etc.) where temperature gradients through the product need to maintained
at a minimum.
When enabled for a loop, MCT will use the lowest process value (if a wait type of Increase is selected) or the
highest process value (if wait type of Decrease is selected) of all the selected loops and/or monitor inputs
programmed for the wait condition, and apply the delta set point offset to that value. The resulting value (lowest
loop or monitor input + Delta SP or highest loop or monitor input
– Delta SP) will be used as the set point for
the loop(s) selected in the list.
To enable delta control, at least one of the available control loops must be selected to have the delta set point
applied and a wait type of Increase or Decrease must be selected. If no loops are selected or a wait type of
Edge is selected, delta control will not be applied and the step will be a normal wait step.
Содержание MCT-CM
Страница 1: ...MCT CM User Manual MCT CM User Manual Rev C August 2022 V1 2eCM30 V1 3eCM22 Supersedes May 2021...
Страница 20: ...MCT CM 20...
Страница 24: ...MCT CM 24...
Страница 45: ...MCT CM 45...
Страница 187: ...MCT CM 187...
Страница 227: ...MCT CM 227 A 4 Control Module Specifications Main CPU...
Страница 228: ...MCT CM 228...
Страница 229: ...MCT CM 229...
Страница 230: ...MCT CM 230...
Страница 231: ...MCT CM 231 Analog I O Card...
Страница 232: ...MCT CM 232...
Страница 233: ...MCT CM 233 Digital Input Card...
Страница 234: ...MCT CM 234...
Страница 235: ...MCT CM 235 Digital Output Card...
Страница 236: ...MCT CM 236...
Страница 237: ...MCT CM 237 Dual Communications Card Barcode and Remote Display Communication Cartridges...
Страница 238: ...MCT CM 238 A 5 Power Supply Specifications...
Страница 239: ...MCT CM 239...
Страница 240: ...MCT CM 240 A 6 Power On Delay Relay Specifications GE1A C10HA110...
Страница 241: ...MCT CM 241...