Invasive Blood Pressure
The “BP* Transducer OFF” message is displayed on the main unit.
Cause
:
The BP (one of 1 to 6) transducer is not connected.
Solution
:
Connect the transducer.
The “BP* Zero Required” message is displayed on the main unit.
Cause
:
The BP zero balance has not been performed since the power was turned ON.
Solution
:
Open the three-way valve of the transducer to air and perform zero balance.
The measurement data of the main unit is displayed as “---”.
Cause
:
The BP zero balance has not been performed since the power was turned ON.
Solution
:
Open the three-way valve of the transducer to air and perform zero balance.
BP value and waveform are not displayed properly on the main unit.
Cause
:
Blood pressure line has not been zero balanced.
Solution
:
Open the three-way valve of the transducer to air and perform zero balance.
The measurement data of the main unit is displayed as “xxx”.
Cause
:
The BP value is outside the measurement range.
Solution
:
Perform zero balance again.
The “BP* Disconnected” message is displayed on the main unit.
Cause
:
While blood pressure is being monitored, BP relay cable has been disconnected from
the 2ch BP Conversion Cable.
Solution 1
:
To cease monitoring, press the Alarm Silence key to clear the message and
silence the alarm.
Solution 2
:
To continue monitoring, plug the 2ch BP Conversion Cable in the BP relay cable. This
will clear the message and silence the alarm.
The “BP* Relay Cable Disconnected” message is displayed on the main unit.
Cause
:
While blood pressure is being monitored, the BP relay cable or 2ch BP Conversion
Cable has been disconnected.
Solution 1
:
To cease monitoring, press the Alarm Silence key to clear the message and
silence the alarm.
Solution 2
:
To continue monitoring, plug the BP Relay Cable in the Multiconnector. This will clear
the message and silence the alarm.
Temperature
The “T* Unknown Sensor” message is displayed on the main unit.
Cause 1
:
A YSI-700 is used.
Solution
:
Use a YSI-400 temperature probe for measurement.
Cause 2
:
There is a contact failure of the temperature probe.
Solution
:
Check if the temperature probe is properly inserted.
Содержание HS-8000 Series
Страница 4: ...Blank Page ...
Страница 20: ...xvi Blank Page ...
Страница 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Страница 24: ...1 4 Blank Page ...
Страница 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Страница 34: ...3 4 Blank Page ...
Страница 62: ...4 28 Blank Page ...
Страница 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Страница 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Страница 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Страница 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Страница 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Страница 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Страница 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Страница 96: ...8 18 Blank Page ...
Страница 132: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100640 201108 ...