Precautions about the Maintenance
Safety Inspection and Maintenance
For safe operation of the equipment, regular inspection and maintenance is required. Once a year,
check all cables, devices, and accessories for damage, earth impedance, earth and leakage
currents, and all alarm functions. Also, ensure that all safety labels are legible. Maintain a record of
these safety inspections.
Immediate maintenance has to be carried out if ;
y
the equipment was subjected to extreme mechanical stress, e.g. after a heavy fall.
y
the equipment was subjected to liquid spill.
y
the monitoring function is interrupted or disturbed.
y
parts of the equipment enclosure are cracked, removed, or lost.
y
any connector or cable shows signs of deterioration.
Reference
Refer to “12. Periodic Check” for details.
W A R N I N G
Never open the housing while the equipment is in operation or connected to
hospital grade outlet as it may result in electric shock.
Maintenance, Modifications, and Repairs
Fukuda Denshi is liable for the safety, reliability, and performance of its equipment only if;
y
Maintenance, modifications, and repairs are carried out by authorized personnel.
y
Components are used in accordance with Fukuda Denshi operating instructions.
Precautions about the Pacemaker
W A R N I N G
z
Minute ventilation rate-adaptive implantable pacemakers can occasionally
interact with the bedside monitor including this unit and diagnostic
equipment, causing the pacemakers to pace at their maximum
programmed rate. The cardiac monitoring and diagnostic equipment may
possibly send wrong information.
If such event occurs, please disconnect the cardiac monitoring and
diagnostic equipment, or follow the procedures described in the operation
manual of the pacemaker.
Reference “Minute Ventilation Rate-Adaptive Pacemakers”
FDA alerts health professionals that minute ventilation rate-adaptive
implantable pacemakers can occasionally interact with certain cardiac
monitoring and diagnostic equipment, causing pacemakers to pace at their
maximum programmed rate.
[October 14, 1998 (Letter: www.fda.gov/cdrh/safety.html) – FDA]
z
Rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely
upon rate meter alarms. Keep pacemaker patients under close
surveillance. See this manual for disclosure of the pacemaker pulse
rejection capability of this equipment.
iv
Содержание HS-8000 Series
Страница 4: ...Blank Page ...
Страница 20: ...xvi Blank Page ...
Страница 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Страница 24: ...1 4 Blank Page ...
Страница 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Страница 34: ...3 4 Blank Page ...
Страница 62: ...4 28 Blank Page ...
Страница 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Страница 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Страница 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Страница 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Страница 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Страница 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Страница 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Страница 96: ...8 18 Blank Page ...
Страница 132: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100640 201108 ...