![Fukuda Denshi HS-8000 Series Скачать руководство пользователя страница 128](http://html.mh-extra.com/html/fukuda-denshi/hs-8000-series/hs-8000-series_service-manual_2344866128.webp)
No. Check Item
Check Procedure
Criteria
10. Cardiac Output (CO)
01
Measurement error
Using a Waveform or CO Simulator,
input 5 L/min or 15 L/min and measure
the error.
Should be within ±10 % of
0.1 to 10 L/min.
Should be within ±15% of
10 to 20 L/min.
Connect a Waveform or CO Simulator,
input blood temperature 36
°
C, 37
°
C,
and 38
°
C, and measure the error.
Blood Temperature should
be within ±0.3
°
C between -1
to 45
°
C.
02 Temperature
measurement
Connect a Waveform or CO Simulator,
input blood temperature -1
°
C and
35
°
C, and measure the error.
Injectate Temperature
should be within ±0.5
°
C
between -1 to 35
°
C.
03
CO Probe off
detection
Check the display by connecting and
disconnecting the probe.
Numeric should disappear.
No. Check Item
Check Procedure
Criteria
11. External Output (Analog Output)
01
ECG output accuracy Input ECG using a comprehensive
tester.
Connect oscilloscope to output and
observe the waveform.
Should be within ±10% of
output 1V to input 1mV.
Should be within ±0.1V to
input 0mV.
02
BP output accuracy
Input BP using a comprehensive
tester.
Connect oscilloscope to output and
observe the waveform.
Should be within ±10% of
output 1V to input
100mmHg.
Should be within ±0.1mV to
input 0mmHg.
No Check Item
Calibration
Reference
12. Calibration
01 Analog
When the measurement error of ECG
or BP is great, calibrate individually.
(Exclusive signal generator for
calibration is required.
)
According to the device,
calibration value of each analog
board is stored in each board.
Therefore, when each or all ECG
or Multi AMP boards are replaced,
make sure to incorporate the
boards into the unit, calibrate
them, and write
calibration/compensation value on
the CPU board.
10
-
10
Содержание HS-8000 Series
Страница 4: ...Blank Page ...
Страница 20: ...xvi Blank Page ...
Страница 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Страница 24: ...1 4 Blank Page ...
Страница 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Страница 34: ...3 4 Blank Page ...
Страница 62: ...4 28 Blank Page ...
Страница 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Страница 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Страница 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Страница 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Страница 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Страница 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Страница 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Страница 96: ...8 18 Blank Page ...
Страница 132: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100640 201108 ...