No
Check Item
Check Procedure
Criteria
13. Electrical Safety
01 Patient
Leakage
Current-I (NC)
Measure the patient leakage current-I
in normal condition using a leak
measurement safety tester. Connect
with the DSC-8500 and measure.
According to test procedure of IEC 60601-1 19.
【
AC/DC
】
Patient Leakage Current-I
(NC)
≤
0.01mA
02 Patient
Leakage
Current-I (SFC)
Measure the patient leakage current-I
in single failure condition using a leak
measurement safety tester. Connect
with the DSC-8500 and measure.
According to test procedure of IEC 60601-1 19.
【
AC/DC
】
Patient Leakage Current-I
(SFC)
≤
0.05mA
03 Patient
Leakage
Current-III (SFC)
Measure the patient leakage current-III
in single failure condition using a leak
measurement safety tester. Connect
with the DSC-8500 and measure.
According to test procedure of IEC 60601-1 19.
Patient Leakage Current-III
(SFC)
≤
0.05mA
04 Patient Current (NC)
Measure normal patient current (NC)
using a leak measurement safety
tester. Connect with the DSC-8500 and
measure.
According to test procedure of IEC 60601-1 19.
Patient Current (NC)
≤
0.01mA
05 Patient Current (SFC) Measure single failure patient current
(SFC) using a leak measurement
safety tester. Connect with the
DSC-8500 and measure.
According to test procedure of IEC 60601-1 19.
Patient Current (SFC)
≤
0.05mA
No Check Item
Check Procedure
Criteria
13. Electrical Safety (*)
Perform the following check item as appropriate. Check these items when you have
disassembled the equipment to check or replace the boards inside the unit or the unit.
06 Withstand
Voltage
Test (A-a2)
Apply AC3000V for 1 minute between
primary power source and exterior
without protective ground (A-a2).
Connect with the DSC-98550 and
measure.
According to test procedure of IEC 60601-1 20.
Should resist the applied
voltage.
07 Withstand
Voltage
Test (B-a)
Apply AC3000V for 1 minute between
primary power source and applied part
(B-a).
According to test procedure of IEC 60601-1 20.
Should resist the applied
voltage.
08 Withstand
Voltage
Test (B-d)
Apply AC1500V for 1 minute between
the exterior and applied part (B-d).
According to test procedure of IEC 60601-1 20.
Should resist the applied
voltage.
10
-
11
Содержание HS-8000 Series
Страница 4: ...Blank Page ...
Страница 20: ...xvi Blank Page ...
Страница 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Страница 24: ...1 4 Blank Page ...
Страница 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Страница 34: ...3 4 Blank Page ...
Страница 62: ...4 28 Blank Page ...
Страница 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Страница 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Страница 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Страница 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Страница 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Страница 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Страница 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Страница 96: ...8 18 Blank Page ...
Страница 132: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100640 201108 ...