9
-
22
9. Remove the “NIBP holder” and “Silicone
Tube” and replace the NIBP unit.
Re-assemble the unit in the order of 9 to 1.
*Pulling the Silicone Tube strongly may cause
a crack. If a crack is found, replace the
Silicone Tube with a new one.
Part Number: 6G0102230 Silicone Tube
C A U T I O N
Make sure that the air connector is in a stress-free situation. If the air connector is under
stress, BP cannot be correctly measured since the connector is loose and the air leaks.
10.Turn ON the power with the Rotary SW of the DSC-8500 to “C”, then press the Menu
→
Maintenance
→
Usage Time keys, press the RESET key in the NIBP Meas. Frequency on the
Usage Time screen, and reset the NIBP Measurement Frequency. And then, return the Rotary SW of the
DSC-8500 to “0”.
11. After the power is turned ON, check that initialization of the NIBP unit is properly completed.
Содержание HS-8000 Series
Страница 4: ...Blank Page ...
Страница 20: ...xvi Blank Page ...
Страница 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Страница 24: ...1 4 Blank Page ...
Страница 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Страница 34: ...3 4 Blank Page ...
Страница 62: ...4 28 Blank Page ...
Страница 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Страница 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Страница 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Страница 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Страница 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Страница 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Страница 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Страница 96: ...8 18 Blank Page ...
Страница 132: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100640 201108 ...