![Fukuda Denshi HS-8000 Series Скачать руководство пользователя страница 26](http://html.mh-extra.com/html/fukuda-denshi/hs-8000-series/hs-8000-series_service-manual_2344866026.webp)
Specification / Performance
This chapter states the specification and performance of this equipment.
Specification
Dimensions
85(W)mm × 100(H)mm × 200(D)mm (not including the protrusion.)
Weight
1.2 kg (not including the accessory)
Environmental Conditions
Operational Temperature
:
10 to 40
°
C
Operational Humidity
:
30 to 85% (non-condensing)
Atmospheric Pressure
:
700 to 1060hPa
Transport / Storage Temperature
:
-10 to 60
°
C
Transport / Storage Humidity
:
10 to 95% (at 60
°
C)
Transport / Storage Atmospheric Pressure
:
700 to 1060hPa
Safety
General Standard (CE):
EN 60601-1:1990+A1:1993+A2:1995
(Medical Electrical Equipment- Part 1: General Requirements for Safety)
EN 60601-1-1:2001
(Medical Electrical Equipment- Part 1-1: General Requirements for Safety
- Collateral standard: Safety Requirements for Medical Electrical
Systems)
EMC Standard (CE) : EN 60601-1-2 Ed. 3.0:2007
(Medical electrical equipment - Part 1-2: General requirements for basic
safety and essential performance Collateral standard:
Electromagnetic compatibility – Requirements and tests)
General Standard :IEC 60601-1:1998+A1:1991+A2:1995
(Medical Electrical Equipment- Part 1: General Requirements for Safety)
IEC 60601-1-1:2000
(Medical Electrical Equipment- Part 1-1: General Requirements for Safety
- Collateral standard: Safety Requirements for Medical Electrical
Systems)
EMC Standard :IEC 60601-1-2 Ed. 3.0:2007
(Medical electrical equipment - Part 1-2: General requirements for basic safety and essential
performance Collateral standard: Electromagnetic compatibility – Requirements
and tests)
Classification
Class of protection against electric shock : Class I equipment
Degree of protection against electric shock : ECG Connector
:
Type CF Applied Part
Multiconnector
:
Type CF Applied Part
NIBP Connector
:
Type BF Applied Part
SpO
2
Connector
:
Type CF Applied Part
Waterproof Level
: IPX0 (no protection)
Sterilization and Disinfection
: Only Cleaning
Usage in Presence of Flammable Gas : Equipment inappropriate to use in presence of
air/flammable anesthetics, or oxygen or nitrous
2
-
2
Содержание HS-8000 Series
Страница 4: ...Blank Page ...
Страница 20: ...xvi Blank Page ...
Страница 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Страница 24: ...1 4 Blank Page ...
Страница 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Страница 34: ...3 4 Blank Page ...
Страница 62: ...4 28 Blank Page ...
Страница 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Страница 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Страница 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Страница 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Страница 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Страница 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Страница 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Страница 96: ...8 18 Blank Page ...
Страница 132: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100640 201108 ...