Cause 1
:
According to the patient condition, the blood pressure may not be correctly measured.
Solution
:
Check the patient condition, and measure again.
Cause 2
:
The cuff application has become loose.
Solution
:
Check if the cuff size matches the patient, and then measure again by wrapping the
cuff properly.
C02-01
When quick measurement, the data cannot be measured during exhaust.
Cause 1
:
According to the patient condition, the blood pressure may not be correctly measured.
Solution
:
Check the patient condition, set the quick measurement of the main unit to OFF, and
measure.
Cause 2
:
The cuff application has become loose.
Solution
:
Check if the cuff size matches the patient, and then measure again by wrapping the
cuff properly.
C02-02
During measurement, the air hose is disconnected from the NIBP Connector.
Cause
:
During measurement, the air hose is disconnected from the NIBP Connector.
Solution
:
Connect the air hose to the NIBP Connector, and then measure.
C03-xx
Exhaust may stop, or the target deflation speed was not achieved.
Cause 1
:
During measurement, an artifact such as body motion may have interfered.
Solution
:
Keep the patient still without body motion as much as possible and measure. For
operation, avoid the artifact by the operator and measure.
Cause 2
:
The air hose is clogged after it is bent or pressed during measurement.
Solution
:
Check the air hose for any bend or press, and then measure.
C04-xx
Inflation for the patient blood pressure is insufficient.
Cause
:
Blood pressure significantly increases from the previous measurement.
Solution
:
Check the cuff application or size, perform manual measurement.
C05-xx
C06-xx
The pulse signal detected during measurement is unstable.
Cause 1
:
During measurement, there is tremble or body motion.
Solution
:
Keep the patient still without tremble or body motion as much as possible and
measure.
Cause 2
:
Arrhythmia event has occurred during the measurement.
Solution
:
If arrhythmia occurs many times, correct measurement cannot be performed.
Measure while arrhythmia occurs as little as possible.
C07-00
The measurement time has exceeded the allowed time.
Cause
:
Measurement is automatically repeated because body motion has been detected or
inflation is insufficient.
Solution
:
Check the cuff application or size, measure by keeping the patient sill without body
motion.
C08-00
Measured PR value is abnormal.
Cause
:
There is tremble or body motion.
Solution
:
Keep the patient still without tremble or body motion as much as possible and
measure.
C09-00
The inflation value has exceeded allowed maximum value.
Cause
:
The cuff is pressed.
Solution
:
Check that the cuff is not pressed, and measure with the patient arms extended as
much as possible.
Содержание HS-8000 Series
Страница 4: ...Blank Page ...
Страница 20: ...xvi Blank Page ...
Страница 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Страница 24: ...1 4 Blank Page ...
Страница 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Страница 34: ...3 4 Blank Page ...
Страница 62: ...4 28 Blank Page ...
Страница 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Страница 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Страница 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Страница 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Страница 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Страница 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Страница 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Страница 96: ...8 18 Blank Page ...
Страница 132: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100640 201108 ...