8
-
12
Cause 2
:
The sensor is exposed to too much ambient light. The detecting part of the sensor is
not covered appropriately.
Solution 1
:
Turn down or turn off the light.
Solution 2
:
Avoid the sensor from exposure to ambient light.
Solution 3
:
Relocate the sensor position.
The “SpO
2
-*Low Perfusion” message is displayed on the main unit.
Cause
:
The amplitude of the pulse waveform is too low, or the sensor is not positioned
correctly.
Solution
:
Check if the light emitting and light receiving parts of the sensor LED are aligned.
The “SpO
2
-*Pulse Search” message is displayed on the main unit.
Cause
:
The amplitude of the pulse waveform is too low, or the sensor is not positioned
correctly.
Solution
:
Check if the light emitting and light receiving part of the sensor LED are aligned.
The “SpO
2
-* Ambient Light Error” message is displayed on the main unit.
Cause
:
The sensor is exposed to too much ambient light. The detecting part of the sensor is
not covered appropriately.
Solution 1
:
Turn down or turn off the light.
Solution 2
:
Avoid sensor from exposure to ambient light.
Solution 3
:
Relocate the sensor position.
The “SpO
2
-* Noise Interference” message is displayed on the main unit.
Cause
:
External signal or energy is interfering with the measurement.
Solution
:
Remove the external interference.
The “SpO
2
-* Check Sensor Conn.” message is displayed on the main unit.
Cause
:
Unrecognizable sensor is connected.
Solution
:
Replace the sensor.
The “SpO
2
-*Low Signal IQ” message is displayed on the main unit.
Cause
:
There is excessive body motion or the sensor position is not appropriate.
Solution 1
:
Check if the light emitting and light receiving parts of the sensor LED are aligned.
Solution 2
:
Relocate the sensor to which the body motion will have less influence.
The “HS-8000 SpO
2
Failure” message is displayed on the main unit.
Cause
:
There is a communication failure with the SpO
2
measurement unit.
Solution
:
A defective cable or SpO
2
measurement unit failure can be considered.
The “SpO
2
-* Disconnected” message is displayed on the main unit.
Cause
:
The SpO
2
relay cable is disconnected during monitoring the SpO
2
.
Solution 1
:
To cease monitoring, press the Alarm Silence key to clear the message and
silence the alarm.
Solution 2
:
To continue monitoring, plug in the SpO
2
relay cable. This will clear the message and
silence the alarm.
The “SpO
2
-* Replace Cable” or “SpO
2
-* Check Cable” message is displayed on the main
unit.
Cause
:
A wrong patient cable is used.
Solution
:
Replace the patient cable. For details of the usable patient cable, refer to “6.
Accessories/Optional Accessories” in the HS-800 series Super Unit Operation
Manual.
Содержание HS-8000 Series
Страница 4: ...Blank Page ...
Страница 20: ...xvi Blank Page ...
Страница 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Страница 24: ...1 4 Blank Page ...
Страница 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Страница 34: ...3 4 Blank Page ...
Страница 62: ...4 28 Blank Page ...
Страница 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Страница 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Страница 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Страница 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Страница 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Страница 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Страница 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Страница 96: ...8 18 Blank Page ...
Страница 132: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100640 201108 ...