55
Management Menu
3-3. Other Option
Sets other parameters other than what are in [Power On Setting] and [Menu Lock
Setting] menus.
Parameter
Descriptions
System Settings
Password
Changes the password to access the [Power On Option], [Menu
Lock], [Other Option], [Password Lock from] menus.
Maximum Characters : 8
At the time of shipment from factory, the password is set at
“0”[zero].
If you have forgotten your password, contact the nearest sales
agency.
Buzzer Volume
Sets the sound volume of the buzzer.
Electrode
Electrode Caution
Electrode
Warning
After 1,000 splices have been made, the caution message prompting
for electrode replacement is displayed when the power is turned on.
When the number of splices exceeds 2,000, the caution message
changes to the warning message. The number of arc discharges at
which to start these messages is set in this field.
Maintenance
Last Maintenance
Next Maintenance
The date of last maintenance and the scheduled date for next
maintenance are entered in the respective fields. This information is
displayed on the [Maintenance Info.] screen. Refer to the next
section.
Arc Calibration
Cleave Limit
Sets the threshold of cleave angle error in the [Arc Calibration].
Max. Num. of
Tests
Sets the number of tests to finish the [Arc Calibration] with “Test
Finished” message.