Maintenance Menu
The splicer has the ability to perform routine maintenance. This section describes how to
use the maintenance menu.
1. Replace Electrodes
Electrodes wear with use and also must be cleaned periodically due to silica buildup. It
is recommended that the electrodes be replaced after 1,000 arc discharges. When the
number of arc discharges reaches a count of 1,000, a message prompting replacement
of the electrodes is displayed immediately after turning on the power. Continuing to
use the electrodes will result in greater splice loss and reduced splice strength.
Replacement Procedure
(1) Execute [Replace Electrode] in
[Maintenance Menu 1].
(2) Instruction messages will appear
on the screen to turn off the power.
Press and hold until the LCD
turn off.
(3) Remove the old electrodes.
Splice Menu
Management Menu