64
Error Message List
Error
Code Error Message
Reason
Solution
01 L-Mismatch
Fiber Count
10 R-Mismatch
Fiber Count
• Mis-selection of splice
mode (number of fibers)
• Select the appropriate splice
mode.
11 LR-Mismatch
Fiber Count
08
-
Mismatch
Fiber Count
(AUTO mode)
• Some fibers are broken.
• Prepare the fiber (stripping,
cleaning, and cleaving) again.
01 Strong
Arc Power
Unable to calibrate due to
strong arc.
02 Weak
Arc Power
Unable to calibrate due to
weak arc.
03 Too Left
Arc
Unable to calibrate due to
poor arc field position (too
far left).
09
04 Too Right
Arc
Unable to calibrate due to
poor arc field position (too
far right).
Replace the electrodes by using
the [Electrode Replace] function.
If this does not eliminate the
problem, consult your nearest
sales agency.
The fiber stuff amount is
insufficient.
If using the standard mode,
check the [Overlap] setting in the
splice mode.
10 Fiber
Separation
The pre-fuse power or
pre-fuse time is set too high.
Check the [Prefuse Power] and
[Prefuse Time] settings in the
splice mode.
• The fiber is not set
correctly at the bottom of
the V-groove.
• Press , and re-position the
fiber again to seat it correctly at
the bottom of the V-groove.
01
• Large fiber offset
• Set the fibers again, or clean the
V-groove.
• The cleave length (bare
fiber) is too short.
• Confirm the setting position of
the stripped fiber end on the
fiber cleaver. Check the cleave
length.
11
02
Bad Fiber
Position
• The stripped fiber end is
not proper position on the
fiber holder.
• Place the fiber to proper
position on the fiber holder.
Содержание KSP75-FP-005468
Страница 2: ......
Страница 9: ...5 Warning and Caution BLANK PAGE ...