59
Maintenance Menu
6. Diagnostic Test Function
The
Splice
Mat
e
TM
has a built in diagnostic test feature that allows the operator to
perform a simple one step evaluation of splicer performance covering several different
critical variables. Perform this function in the event of trouble with splicer operation.
Operation Procedure
(1) The following checks will be made.
Check Item
Description
1 LED Check
Measures and adjusts the brightness of the
illumination LED.
2 Motor Check
Check the Motor Limit Sensor.
3 Dust Check
Checks the optical path for dust or dirt and judges
whether they disturb fiber observation. If
contamination exists, this function indicates the
location.
4 Motor Calibration Automatically calibrates the motor speed.
5 Arc Calibration
Automatically calibrates the arc power factor and
fiber splicing position.
6 I/O Port Check
Checks for normal operation of the input and output
terminals of the internal circuit.
7 Memory Check
Checks the memory of the internal circuit.
(2) Upon completion of all checks and adjustments, a list of results is displayed. If the
splicer fails the dust check, clean the objective lenses. If cleaning does not
eliminate contamination, there is a possibility that the contamination may have
entered the inside optical path. Please contact your nearest sales agency for
additional instructions.
The dust check functions exist as independent instructions in [Maintenance Menu].
• Before the start of the test, remove the fibers from the splicer.
• When item 3 (Dust Check) is completed, prepare and load the fibers into the
splicer and press .