4.13.6. Calibrating the measured value
Description
Measured value can be calibrated manually.
Zero point and span adjustment can be made.
Settable range
(1) Channel designation : CH1, CH2
* When using 2-pipe measurement, configure
both CH1 and CH2.
* No channel setting is necessary for 1-
path/energy measurement and 2-path
measurement.
(2) Zero point
: -5 to +5 [m/s] in terms of flow velocity in piping.
(3) Span
:
±200%
The output value (reading, analog output and total output) is computed by the following expression.
Output =
Measured value × [Span set value %]
100
+ Zero point
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.5.1.)
Operation
(example)
Compensate the zero point to 0.5m
3
/h, and the span by +1%.
* Below is an example for 1-path/energy measurement and 2-path measurement. Refer to Section 4.8.1 for channel
designation of 2-pipe measurement.
Key operation
Description
Display
▼
Press the
key twice to display “OUTPUT SETUP”.
OUTPUT SETUP
ENT
▼
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
▼
Press the
key for 12 times to display “CALIBRATION ZERO”.
CALIBRATION ZERO
0.000 m3/h
ENT
▼
Press the
ENT
key once to blink the cursor.
CALIBRATION ZERO
0
0000.000 m3/h
▼
Press the
key for 6 times to move the cursor.
CALIBRATION ZERO
00000.
0
00 m3/h
▼
Press the
key for 5 times to set “5”.
CALIBRATION ZERO
00000.
5
00 m3/h
Press the
ENT
key once to register.
CALIBRATION ZERO
COMPLETE
↓
ENT
▼
▼
▼
▼
―――
“CALIBRATION ZERO” has been registered.
―――
CALIBRATION ZERO
0.500 m3/h
▼
Press the
key once to display “CALIBRATION SPAN”.
CALIBRATION SPAN
100.0 %
ENT
▼
Press the
ENT
key once to blink the cursor.
CALIBRATION SPAN
1
00.0 %
▼
Press the
key twice to move the cursor.
CALIBRATION SPAN
10
0
.0 %
▼
Press the
key once to set “1”.
CALIBRATION SPAN
10
1
.0 %
Press the
ENT
key once to register.
CALIBRATION SPAN
COMPLETE
↓
ENT
▼
▼
▼
▼
―――
“CALIBRATION SPAN” has been registered.
―――
CALIBRATION SPAN
101.0 %
ESC
Press the
ESC
key once and then press the
key for 3 times to enter
the measurement mode.
0.000 %
0.000 m3
O
u
tp
u
t
0
Zero adj
u
stment mo
v
ement
O
u
tp
u
t
0
Span mo
v
ement
100%
Flo
w
rate
Flo
w
rate
Содержание Time Delta-C FSV-2
Страница 1: ......
Страница 131: ...124 7 APPENDIX 7 1 Specifications...
Страница 132: ...125...
Страница 133: ...126...
Страница 149: ......