4.12.4. LCD indication
Description
Selecting the source channel
Specify of which channel (path 1, path 2, or calculated value) values are to be displayed in the first and the second line.
Channel selection: CH1, CH2,CH3
<Note> In 1-path/energy measurement, no parameters need to be specified by channel number
Flow velocity indication
Selectable flow velocity units: m/s (if SYSTEM UNIT was set to METRIC) (See section 4.11.1)
<Note> The decimal point position is fixed. (3 digits after a decimal point)
Flow rate indication
Selectable flow rate indications: Actual value reading, % reading.
<Note> The indication unit is as selected by FLOW UNIT. (See section 4.11.2.)
Total flow indication
Selectable total indications: Actual total value reading (forward/reverse flow), total pulse count (forward/reverse flow).
<Note> The indication unit is as selected by TOTAL UNIT. (See section 4.11.3.)
* The following items are used in 1-path/energy measurement.
Instantaneous energy indication
Selection: Actual value reading, % reading
<Note> The indication unit is as selected by THERMAL UNIT. (See section 4.11.5)
Total energy indication
Selection: Actual total value reading (heating/cooling), energy total pulse count (heating/cooling).
<Note> The indication unit is as selected by TOTAL(THERMAL). (See section 4.11.6)
Temperature indication
Selection: supply temperature, return temperature, temperature difference
<Note> The indication unit is as selected by TEMPERATURE. (See section 4.11.4.)
<Note> The decimal point position is fixed. (
℃
: 1 digit after a decimal point, K: 2 digits after a decimal point)
How to set the indication
Set the DISPLAY setting mode to 1st ROW (for indication on 1st line) or 2nd ROW (for indication on 2nd line), and further select
indication contents.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.5.1.)
Operation
(example)
Display the 1st line of LCD indication in percentages (%).
* Below is an example for 1-path/energy measurement. Refer to Section 4.8.1 for channel designation of 2-path
measurement and 2-pipe measurement.
Key operation
Description
Display
▼
Press the
key twice to display “OUTPUT SETUP”.
OUTPUT SETUP
ENT
▼
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
▼
Press the
key for 3 times to display “DISPLAY”.
DISPLAY
1ST.ROW
ENT
▼
Press the
ENT
key once to blink the cursor.
DISPLAY
1ST.ROW
ENT
▼
Press the
ENT
key again, and select “1ST. LOW”.
1ST. LOW
VELOCITY
▼
Press the
key twice to display “FLOW RATE(%)”.
1ST. LOW
FLOW RATE(%)
ENT
▼
Press the
ENT
key once, and select and fix “FLOW RATE(%)” to display
“1:DECIMAL POINT”.
1:DECIMAL POINT
.
▼
Press the
key once to shift the decimal point position to next place.
1:DECIMAL POINT
.
Press the
ENT
key once to register.
1:DECIMAL POINT
COMPLETE
↓
ENT
▼
▼
▼
▼
―――
FLOW RATE(%) indication has been set.
―――
1:DECIMAL POINT
.
ESC
Press the
ESC
key for 3 times and then the
key for 3 times to enter
the measurement mode.
0.00 %
0.000 m3