-89-
4.15.3. Constant current output
Description
Generates a fixed value output of analog signal.
Application example: The operation of a connected receiver is checked by generating a fixed value output of analog signal.
In the constant current setting mode (OUTPUT SETTING), set the constant current output value.
Settable range: -20%(0.8mA) to +120%(23.2mA)
This operation sets AO1 and AO2 the same contact action.
Before operation, check whether it is permitted to change AO output.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.5.1.)
Operation
(example)
Set the constant current output of 50% (12mA).
Key operation
Description
Display
▼
Press the
key for 4 times to display “MAINTENANCE MODE”.
MAINTENANCE MODE
ENT
▼
Press the
ENT
key once to display “RAS INFORMATION”.
RAS INFORMATION
▼
Press the
key for 3 times to display “OUTPUT SETTING”.
CURRENT
OUTPUT SETTING
ENT
▼
Press the
ENT
key once to display the setting screen.
OUTPUT SETTING
0 %
ENT
▼
Press the
ENT
key once to blink the cursor.
Note) Constant current output starts.
OUTPUT SETTING
+
000 %
▼
Enter “5” by using the
key and the
key.
OUTPUT SETTING
+0
5
0 %
Press the
ENT
key once to output 12mA.
OUTPUT SETTING
COMPLETE
↓
ENT
▼
▼
▼
▼
―――
Outputting 12mA.
―――
OUTPUT SETTING
50 %
ESC
▼
Press the
ESC
key once to stop constant current output.
Note) Current output is in the measurement status.
CURRENT
OUTPUT SETTING
ESC
Press the
ESC
key once and then press the
key once to enter the
measurement mode.
0.000 %
0.000 m3