-74-
4.13.5.4. Total switch
Description
Contact output DO1, DO2, DO3, and/or DO4 are activated when the totalized value
exceeds a setting.
Contact output DO1, DO2, DO3, and/or DO4 are activated when the instantaneous
energy exceeds a setting. (for 1-path/energy measurement)
Settable range: 0.000001 to 99999999
Action
:TOTAL SWITCH, TOTAL SWITCH(Thermal)
Contact action:
ACTIVE ON : DO1/DO2/DO3/DO4: Normally off
ACTIVE OFF : DO1/DO2/DO3/DO4: Normally on
Note) Different values can be assigned to DO1, DO2, DO3, and DO4.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.5.1.)
Operation
(example)
Set the DO1 output to “TOTAL SWITCH”, and change the setting value from 10000[m
3
] to 100[m
3
].
Also, set the contact to “ACTIVE ON”.
* Below is an example for 1-path/energy measurement. Refer to Section 4.8.1 for channel designation of 2-path
measurement and 2-pipe measurement.
Key operation
Description
Display
▼
Press the
key twice to display “OUTPUT SETUP”.
OUTPUT SETUP
ENT
▼
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
▼
Press the
key for 8 times to display “DO1 OUT”.
* Use the
key to select “DO2 OUT”, “DO3 OUT”, or “DO4 OUT”.
DO1 OUT
NOT USED
ENT
▼
Press the
ENT
key once to blink the cursor.
DO1 OUT
NOT USED
▼
Press the
key for 6 times to display “TOTAL SWITCH” on the 2nd
line.
DO1 OUT
TOTAL SWITCH
ENT
▼
Press the
ENT
key once to display the setting screen of “TOTAL
SWITCH”.
TOTAL SWITCH
10000 m3
ENT
▼
Press the
ENT
key once to blink the cursor.
TOTAL SWITCH
0
0010000 m3
▼
Press the
key for 3 times to move the cursor.
TOTAL SWITCH
000
1
0000 m3
▼
Press the
key for 10 times to set “0”.
TOTAL SWITCH
000
0
0000 m3
▼
Press the
key twice to move the cursor.
TOTAL SWITCH
00000
0
00 m3
▼
Press the
key once to set “1”.
TOTAL SWITCH
00000
1
00 m3
Press the
ENT
key once to register.
TOTAL SWITCH
COMPLETE
↓
ENT
▼
▼
▼
▼
―――
“TOTAL SWITCH” has been registered.
―――
STATUS OUT
CONTACT ACTION
ENT
▼
Press the
ENT
key once to display “CONTACT ACTION”.
CONTACT ACTION
ACTIVE ON
Press the
ENT
key once to register “ACTIVE ON”(normally off).
* To select normally on, press the
key.
CONTACT ACTION
COMPLETE
↓
ENT
▼
▼
▼
▼
―――
“ACTIVE ON” has been registered.
―――
STATUS OUT
CONTACT ACTION
ESC
Press the
ESC
key twice and then press the
key for 3 times to enter
the measurement mode.
0.000 %
0.000 m
3
Time
Setting
v
al
u
e
Total
v
al
u
e
O
N
or OFF