54
The wiring and the access to do so, can be dif
fi
cult according the position of the transmitter. The
electronics housing position can be modi
fi
ed by turning it by steps of 90 or 180 degrees.
The electronics housing is
fi
xed by hexagonal socket bolts (M6 x 12). Loosen the bolts and turn
the electronics housing to the right or left hand side; fasten the
fi
xing screws after this operation
again
The analog or digital indicator can be turned ±180
˚
in 90
˚
increments because it is connected
with a pin plug.
If the assembly parts must be turned over than 90°, or if the
position is already amended since the delevery by Fuji, it's
necessary to remove the electronic unit from the housing
and disconnect the flat cable from the electronic measur-
ing cell before turn the housing.
If the transmission unit has been turned excessively without
removing the electronics unit, straighten the
fl
at cable which
connects the electronics unit in the transmission unit and the
detecting unit, and set the transmission unit again.
%
0
20
40
60
80
100
%
0
20
40
60
80
100
%
0
20
40
60
80
10
0
Hex
socket bolt
Change of electronics housing position
DANGER
PRECAUTION
DANGER
Avoid the following procedure in an explosionproof area.
Avoid the following procedure in an explosionproof area.
Change of indicator angle