5
EMC Directive (20
1
4/
30
/E
U
)
All models of
FCX
series
electronic pressure
transmitters type
FCX-AII & CII
are in accordance with :
• The harmonized standard EN 61326-1 : 20
13
(Electrical equipment for measurement, control and laboratory use
-
EMC requirements
- Part 1: General requirments
).
Emission limits :
EN 61326-1 : 20
13
Frequency range (MHz)
Limits
Basic standard
30 to 230
40 dB (
μ
V/m) quasi peack, measured at 10m distance
EN 55011 / CISPR 11
Group 1 Class A
230 to 1000
47 dB (
μ
V/m) quasi peack, measured at 10m distance
Minimum
requirements
for Immunity tests
:
EN 61326-1 : 20
13
(Table 2)
Phenomenon
Test value
Basic standard
Performance criteria
Electrostatic discharge (EDS)
4 kV (Contact)
EN
/IEC
61000-4-2
B
8 kV (Air)
Electromagnetic
fi
eld
EN/IEC
61000-4-3
A
10
V/m (80 to 1000 MHz)
3 V/m (1.4 to 2.0 GHz) 1
V/m (2.0 to 2.7 GHz)
Rated power frequency
30 A/m
(50 Hz, 60 Hz)
EN/IEC
61000-4-8
A
Magnetic
fi
eld
Burst
2 kV (5/50
ns
, 5 kHz
EN/IEC
61000-4-4
B
Surge
EN/IEC
61000-4-5
B
1 kV
-
Line to line
2 kV
-
Line to
ground
Conducted RF
disturbances
3 V
rms
(150 kHz
-
80 MHz)
EN/IEC
61000-4-6
A
Performance criteria
(A & B)
:
According EN 61326-1 :
2
013