76
Level pressure range
FKE model
kPa (mbar)
0,32~32 (3,2~320)
1,3~130 (13~1300)
5~500 (50~5000)
Pressure range with remote seal
FKB model
kPa (bar)
1,3~130 (0,013~1,3)
5~500 (0,05~5)
30~3000 (0,3~30)
100~10000 (1~100)
500~50000 (5~500)
Differential pressure range with remote seal(s)
FKD model
kPa (mbar)
0,32~32 (3,2~320)
1,3~130 (13~1300)
5~500 (50~5000)
CALIBRATION
A2
Preparation for calibration
The transmitter should be calibrated in a calibration room. For calibration of each transmitter,
the following devices are required.
• Pressure source and pressure measuring equipment (should have as high an accuracy as
possible 0,05% mini)
* Measurable ranges are listed in the table below.
• Power supply : DC power supply (24 V DC) or Fuji Electric FCX series power supply unit
• Load resistor : Standard resistor 250
Ω
(within ±0.0125
Ω
)
• Measuring device: Digital voltmeter (capable of measuring transmitter output with an accu-
racy better than 0.1%)
* Use meter having a 5-digit display.
• Hand Held Communicator (HHC) type FXW
Differential pressure range
FKC model
kPa (mbar)
0,1~1 (1~10)
0,1~6 (1~60)
0,32~32 (3,2~320)
1,3~130 (13~1300)
5~500 (50~5000)
30~3000 (300~30000)
Pressure range
FKG model
kPa (bar)
1,3~130 (0,013~1,3)
5~500 (0,05~5)
30~3000 (0,3~30)
100~10000 (1~100)
500~50000 (5~500)
Absolute range
FKA model
kPa abs. (bar abs.)
1,6~16 (0,016~0,16)
1,6~130 (0,016~1,3)
5~500 (0,05~5)
30~3000 (0,3~30)
Measurable range