3.2 Operation
(1) Operation of gauge and pressure transmitter :
Open the valve slowly to apply a
pressure.
When a pressure is applied, the trans-
mitter is set inthe operating status.
(2) Operation of differential pressure transmitter :
Set the operating status by manipulating the manifold valve.
Open
Open
Stop valve on the
LP side
Stop valve on the
HP side
Equalizing
valve
Close
Make sure the equalizing valve is open
Open the stop valve on the HP side
slowly.
Close the equalizing valve.
Open