50
5.4 Adjustment after replacement of the amplifier unit or the measuring cell
Adjustment
After completion of the assembly work mentioned above, use the following procedures for ad-
justment and setting. Adjustment should be performed using the HHC communicator.
(1) After replacement of electronics unit (including replacement of internal parts)
Step
Adjustment
item
Page
1
Constant current output (output circuit)
29-30
Digital/Analog
2
TAG
N°
19
3
Type of transmitter
19
4
Unit
20
5
Adjustment of the zero and span
21
6
Calibration of the zero and span adjustment
25
7
Damping
22
8
Output mode LIN /
23
9
Burnout
24
10
Lock of adjustment local function
28
(2) After replacement of detecting unit (including replacement of internal parts)
Step
Adjustment
item
Page
1
Adjustment of the zero and span
21